- Human HPV-IgG ELISA Kit Human HPV-IgG ELISA Kit
Human HPV-IgG ELISA Kit
- Description
- Expédition et retours
- Avis
The HPV IgG ELISA Kit is designed for human papillomavirus antibody detection in research and epidemiological studies. Suitable for virology and immunology laboratories.
GRSC Store supplies HPV ELISA kits with fast delivery, COA documentation, and technical support.
HPV IgG ELISA Kit FAQ (Human Papillomavirus IgG ELISA Kit)
Q: What is the HPV IgG ELISA Kit used for?
A: The HPV IgG ELISA Kit is used for the detection and quantitative measurement of Human Papillomavirus (HPV) IgG antibodies in research samples to support virology, immunology, and infectious disease studies.
Q: Which sample types are compatible with the HPV IgG ELISA Kit?
A: Most HPV IgG ELISA Kits support serum, plasma, and other human laboratory research samples depending on the kit protocol and specifications.
Q: What research applications commonly use HPV IgG ELISA Kits?
A: HPV IgG ELISA Kits are widely used in HPV infection research, vaccine response studies, immune monitoring, epidemiological investigations, cervical cancer research, and antibody response analysis.
Q: Is this kit suitable for academic and biomedical laboratories?
A: Yes. HPV IgG ELISA Kits are commonly used in universities, biomedical research institutes, hospitals, diagnostic research centers, and pharmaceutical research laboratories.
Q: Can GRSC Store support bulk institutional orders?
A: Yes. GRSC Store supports bulk procurement for laboratories, universities, hospitals, biotechnology companies, and healthcare research organizations.
Q: Are technical manuals and datasheets available with the kit?
A: Yes. Most HPV IgG ELISA Kits include detailed protocols, product datasheets, and technical documentation to support laboratory research applications.
Product No.: YP-W12159
Specification: 96T
Kit performance Detection range: 20 IU/mL – 640 IU/mL.
Sensitivity: Minimum detected dose is less than 1.0 IU/mL.
Precision: In-batch variation coefficient CV% is less than 10%; interbatch variation coefficient CV% is less than 15%.
Recovery rate: The recovery rate is between 85% and 115%.
Specificity: This kit recognizes natural and recombinant HumanHPV-IgG(HPV-IgG)without crossing with structural analogs.
Stability: Storage at 2℃-8℃, valid for 6 months.
Usage: Used to detect the concentration of HumanHPV-IgG(HPV-IgG)in samples such as serum, plasma, cell culture supernatant and tissue.
Shelf life: 2℃-8℃, valid for 6 months.
Experimental principle
The kit uses an indirect enzyme-linked immunosorbent assay (ELISA). To the coated microwells precoated with HumanHPV-IgG(HPV-IgG) capture antigen, the sample to be tested and the standard product were added in turn, and then the HRP-labeled detection antibody was added, incubated and thoroughly washed. Substrates A and B are added, and the substrate produces a blue product under HRP catalysis, which is finally converted to yellow under the action of a stop solution (acid solution). The absorbance (OD value) was measured at a wavelength of 450 nm on the microplate reader, and the absorbance (OD value) was positively correlated with the concentration of HumanHPV-IgG(HPV-IgG) in the sample to be tested. By fitting the calibrator curve, the concentration of HumanHPV-IgG(HPV-IgG) in the sample can be calculated.
Kit components and storage Unopened kits are kept at 2-8 degrees and expired kits must not be used.
|
Components |
48 hole configuration |
96 hole configuration |
Store after opening |
|
Pre-coated enzyme plate |
48T |
96T |
2-8℃14 days |
|
Standard |
0.3mL*6 tube |
0.3mL*6 tube |
2-8℃14 days |
|
Sample dilution |
3 ml |
6 ml |
2-8℃180 days |
|
HRP-labeled antibodies |
5 ml |
10 ml |
2-8℃14 days |
|
Color-producing substrate A |
3 ml |
6 ml |
2-8℃180 days |
|
Color-producing substrate B |
3 ml |
6 ml |
2-8℃180 days |
|
Termination fluid |
3 ml |
6 ml |
2-8℃180 days |
|
20× washing solution |
15 ml |
25 ml |
2-8℃180 days |
|
Seal membrane |
2 photos |
2 photos |
|
|
manual |
1 serving |
1 serving |
|
|
Self-sealed bag |
1 |
1 |
|
The calibrator concentrations are: 640、320、160、80、40、20 IU/mL in turn.
Note:
1: Before use, please check whether the label and quantity of the reagent in the kit are consistent with the table.
2: If the components of the kit need to be used again, please make sure that they are not contaminated after the last use. 3: The enzyme label plate has not been used for a single time. Remember to seal it and store it at 2-8℃.
Self-provided testing equipment required for testing
(not provided, but can be assisted in purchasing)
1) Microplate reader that can detect absorbance of 450 nm
2) Pipet and gun tip, sample filling tank
3) 37℃ constant temperature box or water bath pot
4) Test tubes, centrifuge tubes, measuring cylinders, etc. for preparing reagents
5) Distilled water or deionized water
6) Vortex oscillator, microplate oscillator
Notes
1) It is for scientific research only and shall not be used for clinical diagnosis.
2) Used within the validity period marked by the kit, expired products shall not be used.
3) Do not mix with other manufacturers' kits or components, use the sample diluent that is equipped with the kit.
4) If the sample value is higher than the highest standard concentration value, please dilute the sample appropriately before re-measurement.
5) The heterophilic antibodies such as human anti-mouse in the sample to be tested will interfere with the detection results. Please excrete this factor before testing.
6) The test results obtained by other methods are not directly comparable to the test results of this kit.
7) Please wear experimental clothes and latex gloves to protect yourself during the test. Especially when testing blood or other body fluid samples, please follow the National Biological Laboratory Safety Protection Regulations.
8) Incubate strictly in accordance with the prescribed time and temperature to ensure accurate results. All reagents must reach room temperature 20-25°C before use. Refrigerate immediately after use.
9) Incorrect washing of the board can lead to inaccurate results. Make sure to suck the liquid in the pores as much as possible before adding the substrate. Do not let the micropores dry out during incubation.
10) Eliminate residual liquid and finger prints on the bottom of the plate, otherwise it will affect the OD value.
11) The substrate color-producing liquid should be colorless or very light.
12) Avoid cross-contamination of reagents and specimens to avoid causing erroneous results.
13) Avoid direct exposure of strong light during storage and incubation.
14) The microplate reader used for detection needs to be installed with a filter that can detect wavelengths of 450±10nm, with an optical density range between 0-3.5. It is recommended to warm up 15 minutes in advance when using it.
15) The EP tube and suction head used in the test are both used once and are strictly prohibited from being mixed.
Sample preparation and preservation
The following are just general guidelines for sample collection and preservation. During the collection and storage of all samples, sodium azide must not be used as a preservative. If the sample is not analyzed immediately, it should be aliquoted and frozen and avoid repeated freeze-thawing.
Cell culture supernatant - centrifuge to remove the precipitate, analyze immediately or aliquot it and store it frozen at -20°C.
Serum - Collect blood with a clean test tube, coagulate at room temperature for 30 minutes, centrifuge 2000×g for 20 minutes, and collect serum. Immediately analyze or aliquot it and store it frozen at -20°C.
Plasma - Use heparin, citrate or EDTA to anticoagulate, and centrifuge at 2-8°C for 2000×g for 20 minutes within 30 minutes after blood drawing. To eliminate the effects of platelets, it is recommended to further centrifuge at 2-8°C for 10,000×g for 10 minutes. Immediately analyze or aliquot it and cryopreserve at -20°C.
Cell Lysate - For adherent cells, remove the culture medium and wash it with PBS, saline or serum-free culture medium. Add an appropriate amount of lysate and blow it with a gun for a few times to make the lysate and the cells fully contact. Usually after 10 seconds, the cells are lysed. For suspended cells, collect the cells by centrifugation and wash them with PBS, saline or serum-free culture medium. Add an appropriate amount of lysate, blow the cells with a gun, and flick them with your fingers to fully lyse the cells. After sufficient cleavage, centrifuge at 10000-14000×g for 3-5 minutes and remove the supernatant. Immediately analyze or aliquot it and cryopreserve at -20°C.
Tissue homogenate - Rinse the tissue with pre-cooled PBS (0.01M, pH=7.4) to remove residual blood (the lysed red blood cells in the homogenate will affect the measurement results), and cut the tissue after weighing. The sheared tissue and the corresponding volume of PBS (usually at a weight-volume ratio of 1:9, for example, 1g of tissue sample corresponds to 9mL of PBS. The specific volume can be adjusted appropriately according to experimental needs and recorded. It is recommended to add protease inhibitors to PBS) to a glass homogenizer and grind it thoroughly on ice. To further lyse tissue cells, the homogenate can be sonicated or repeatedly freeze-thawed. Finally, the homogenate was centrifuged at 5000×g for 5~10 minutes, and the supernatant was taken for testing.
Urine – collect with a sterile tube and centrifuge at 2000×g for 20 minutes. Collect the supernatant carefully. If precipitation forms, centrifuge again.
Reagent preparation
1. Before use, all components must be re-warmed for at least 60 minutes to ensure full re-warming to room temperature.
2. Concentrated washing liquid: The concentrated washing liquid taken out of the refrigerator will produce crystallization, which is a normal phenomenon. The crystallization is completely dissolved by heating in a water bath. The concentrated washing liquid and distilled water are diluted at 1:20, that is, 1 part of the concentrated washing liquid, and 19 parts of the distilled water are added.
3. Substrate: substrate liquids A and B, mix thoroughly at 1:1 volume before use, and use within 15 minutes after mixing.
Operating procedures: All reagents and components are first restored to room temperature, and are recommended to make compound holes.
1. Prepare the working liquid of various components of the kit according to the method described in the previous instructions.
2. Remove the required slats from the aluminum foil bag, and seal the remaining slats with a self-sealing bag and put them back into the refrigerator.
3. Set up standard product wells and sample wells, and add 50μL of standard products of different concentrations to each standard product well;
4. Add 50μL of the sample to be tested to the sample well; do not add the blank well.
5. Except for the blank wells, add 100μL of horseradish peroxidase (HRP)-labeled detection antibody to each well of the standard well and sample wells, seal the reaction wells with a sealing membrane, and incubate for 60 minutes at 37℃ water bath or constant temperature box.
6. Discard the liquid, pat dry on the absorbent paper, fill each well with washing liquid (350μL), let stand for 1 minute, shake off the washing liquid, pat dry on the absorbent paper, and repeat the washing plate 5 times (you can also use the washing machine to wash the plate).
(Tip: To obtain ideal experimental results, the residual liquid must be completely removed. After washing the plate, please perform the next step immediately and do not let the microplate dry.)
7. Add 50μL of substrate A and B each well, and incubate at 37°C for 15 minutes away from light.
8. Add 50μL of the stop solution per well, and within 15 minutes, determine the OD value of each well at a wavelength of 450nm.
Result calculation
1. Use the concentration of the standard substance as the abscissa and the corresponding absorbance (OD value) as the ordinate. Use computer software and four-parameter Logistic curve fitting (4-pl) to create a standard curve equation. Through the absorbance (OD value) of the sample value), use the equation to calculate the concentration value of the sample. [Calculate using ELISA Calc software. It is recommended to use four-parameter fitting for the standard curve, but it is not the only fitting method]
2. If the sample is diluted, the concentration value measured by the above method must be multiplied by the dilution factor to determine the final value of the sample. concentration. Note: Experimenters need to establish a standard curve based on their own experiments. For each test, a standard curve must be established for each enzyme plate. The following curves are for reference only!
Problem Analysis
If the experimental results are not good, please take pictures of the color development results in time, save the experimental data, keep the used strips and unused reagents, and then contact our company's technical support to solve the problem for you.
Fast Questions and Answers
|
Problem description |
Possible reasons |
Corresponding countermeasures |
|
standard curve gradient difference |
Incorrect liquid aspiration or addition |
Check pipettes and tips |
|
Equilibration time is too short |
Ensure sufficient balancing time |
|
|
Incomplete washing |
Ensure the washing time and number of washes and the amount of liquid added to each hole |
|
|
Very weak or colorless |
Incubation time too short |
Ensure adequate incubation time |
|
The experimental temperature is incorrect |
Use recommended experimental temperatures |
|
|
Insufficient reagent volume or missing addition |
Check the liquid aspiration and addition process to ensure that all reagents are added in sufficient order and in sufficient quantities. |
|
|
Incorrect dilution |
||
|
Enzyme label inactivation or substrate failure |
Mix enzyme conjugate and substrate and check by rapid color development |
|
|
Reading value is low |
Microplate reader settings are incorrect |
Check the wavelength and filter settings on the microplate reader |
|
Turn on the microplate reader and preheat it in advance |
||
|
Large coefficient of variation |
Adding fluid incorrectly |
Check the filling situation |
|
High background value |
The working concentration of the detection antibody is too high |
Use the recommended dilution factor |
|
Incomplete washing of enzyme plate |
Ensure that each step of cleaning is complete; if using an automatic plate washer, please check whether all outlets are blocked; whether the washing solution provided in the kit is used |
|
|
The lotion is contaminated |
Prepare fresh lotion |
|
|
Low sensitivity |
Improper storage of ELISA kits |
Store relevant reagents according to instructions |
|
Not terminated before reading |
Stop solution should be added to each well before OD reading |
Statement
1. Due to the current conditions and scientific and technological level, it is not possible to conduct comprehensive identification and analysis of all raw materials. This product may have certain quality and technical risks.
2. This kit removes/reduces some endogenous interfering factors in biological samples during the development process. Not all possible influencing factors have been removed.
3. The final experimental results are closely related to factors such as the effectiveness of the reagents, the relevant operations of the experimenter, and the experimental environment at the time. Our company is only responsible for the kit itself and is not responsible for the sample consumption caused by the use of the kit. Please use The user should fully consider the possible usage of the sample and reserve sufficient samples before use.
4. In order to achieve good experimental results, please only use the reagents provided in our company's kits, do not mix products from other manufacturers, and operate in strict accordance with the instructions.
5. Due to incorrect reagent preparation and microplate reader parameter settings during the operation, abnormal results may result. Please read the instructions carefully and adjust the instrument before the experiment.
6. Even if operated by the same personnel, different results may be obtained in two independent experiments. In order to ensure the reproducibility of the results, it is necessary to control every step of the experimental process.
7. The kits will undergo strict quality inspection before shipment. However, due to factors such as transportation conditions, differences in experimental equipment, etc., user test results may be inconsistent with factory data.
8. This kit has not been compared with similar kits from other manufacturers or products that detect the same target substance using different methods, so inconsistent test results cannot be ruled out.
9. The kit is for research use only. If it is used for clinical diagnosis or any other purpose, our company will not be responsible for any problems arising therefrom, nor will we assume any legal liability.
Propriété intellectuelle
- Lorem ipsum dolor sit amet, consectetur adipiscing elit. Etiam quis diam erat. Duis velit lectus, posuere a blandit sit amet, tempor at lorem. Donec ultricies, lorem sed ultrices interdum.
- Lion metus luctus sem, vel vulputate diam ipsum sed lorem. Donec tempor arcu nisl, et molestie massa scelerisque ut. Nunc au rutrum Leo. Mauris metus mauris, tristique quis sapien eu, rutrum vulputate enim.
- Mauris tempus erat laoreet turpis lobortis, eu tincidunt erat fermentum.
- Aliquam non tincidunt urna. Entier tincidunt nec nisl vitae ullamcorper. Proin sed ultrices erat. Praesent divers ultrices massa au faucibus.
- Dignissim énéen, orci sed faucibus pharetra, dui mi dignissim tortor, sit amet condimentum mi ligula sit amet augue.
- Pellentesque vitae eros eget enim mollis placerat.
Terminaison
- Il existe de nombreuses variantes de passages du Lorem Ipsum disponibles
- Si vous allez utiliser un passage
- Tous les générateurs Lorem Ipsum sur Internet ont tendance à répéter des morceaux prédéfinis si nécessaire
- Il utilise un dictionnaire de plus de 200 mots latins, combiné à une poignée de structures de phrases modèles
- L'intérêt d'utiliser Lorem Ipsum est qu'il a une distribution de lettres plus ou moins normale
- De nombreux logiciels de publication assistée par ordinateur et éditeurs de pages Web utilisent désormais Lorem Ipsum comme texte modèle par défaut.
Modifications apportées au présent accord
Nous nous réservons le droit, à notre entière discrétion, de modifier ou de remplacer les présentes Conditions générales en publiant les conditions mises à jour sur le Site. Votre utilisation continue du Site après de telles modifications vaut acceptation des nouvelles Conditions générales.
Contactez-nous
Si vous avez des questions concernant cet accord, veuillez nous contacter en remplissant ce formulaire de contact.